CVE-2024-9832
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Scoring
- Severity
- CRITICAL
- CVSS base score
- 9.3
- CVSS vector
- CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- EPSS probability
- 0.18%
- CWE
- CWE-307
- Published
- 2024-11-14
- Last modified
- 2026-03-13
Affected products
- Baxter Life2000 Ventilation System
Weakness type