CVE-2024-9832

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Scoring

Severity
CRITICAL
CVSS base score
9.3
CVSS vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS probability
0.18%
CWE
CWE-307
Published
2024-11-14
Last modified
2026-03-13

Affected products

Weakness type

Related vulnerabilities

Markdown version · Browse all CVEs