# CVE-2024-9832

## Summary

- **CVE ID:** CVE-2024-9832
- **Severity:** CRITICAL
- **CVSS Score:** 9.3 (CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H)
- **CWE:** CWE-307
- **Published:** Nov 14, 2024
- **Last Modified:** Mar 13, 2026

## Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

## Affected Products

- Baxter — Life2000 Ventilation System (06.08.00.00 and prior)

## References

- [CNA](https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01)

## Exploitation Prediction (EPSS)

- **EPSS Score:** 0.18%
- **EPSS Percentile:** 7.3

---
_Exported from OnDuty AI Vulnerability Intelligence on 2026-09-10._