CVE-2024-48970
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.
Scoring
- Severity
- CRITICAL
- CVSS base score
- 9.3
- CVSS vector
- CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- EPSS probability
- 0.21%
- CWE
- CWE-1191
- Published
- 2024-11-14
- Last modified
- 2026-03-13
Affected products
- Baxter Life2000 Ventilation System
Weakness type
Related vulnerabilities
- CVE-2026-15203 — Debug interfaces are accessible by default in Danfoss iC7 Automation SP, iC7 Marine and iC7 7Hybrid software
- CVE-2026-8989 — Open Recovery Mode
- CVE-2026-8988 — Access to Bootloader
- CVE-2024-36319 — Debug code left active in AMD's Video Decoder Engine Firmware (VCN FW) could allow a attacker to...
- CVE-2025-52533 — Improper Access Control in an on-chip debug interface could allow a privileged attacker to enable a...
- CVE-2025-15083 — TOZED ZLT M30s UART on-chip debug and test interface with improper access control
- CVE-2025-36755 — CleverDisplay BlueOne unauthorized BIOS access through physical USB keyboard
- CVE-2025-12114 — Serial Console Enabled