# CVE-2024-48970

## Summary

- **CVE ID:** CVE-2024-48970
- **Severity:** CRITICAL
- **CVSS Score:** 9.3 (CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H)
- **CWE:** CWE-1191
- **Published:** Nov 14, 2024
- **Last Modified:** Mar 13, 2026

## Description

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

## Affected Products

- Baxter — Life2000 Ventilation System (06.08.00.00 and prior)

## References

- [CNA](https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01)

## Exploitation Prediction (EPSS)

- **EPSS Score:** 0.21%
- **EPSS Percentile:** 11.6

---
_Exported from OnDuty AI Vulnerability Intelligence on 2026-09-10._