CVE-2025-0683

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Scoring

Severity
HIGH
CVSS base score
8.2
CVSS vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N
EPSS probability
0.79%
CWE
CWE-359
Published
2025-01-30
Last modified
2026-03-13