CVE-2021-4479

Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

Scoring

Severity
MEDIUM
CVSS base score
6.3
CVSS vector
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
EPSS probability
0.24%
CWE
CWE-1286, CWE-1286
Published
2026-06-02
Last modified
2026-06-10

Affected products

Weakness type

Related vulnerabilities