CVE-2019-25723

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Scoring

Severity
MEDIUM
CVSS base score
6.3
CVSS vector
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
EPSS probability
0.24%
CWE
CWE-1286
Published
2026-06-02
Last modified
2026-06-03

Affected products

Weakness type

Related vulnerabilities

Markdown version · Browse all CVEs