# CVE-2019-25719

## Summary

- **CVE ID:** CVE-2019-25719
- **Severity:** HIGH
- **CVSS Score:** 8.8 (CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N)
- **CWE:** CWE-924
- **Published:** Jun 2, 2026
- **Last Modified:** Jun 3, 2026

## Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

## Affected Products

- Dräger — Infinity Acute Care System (VG4.1.1)
- Dräger — Infinity Acute Care System (VG4.0.3)
- Dräger — Infinity Acute Care System (lower than VG4.0.3)
- Dräger — Infinity Acute Care System (VG4.2)
- Dräger — Standalone Infinity M540 patient monitor (VG4.1.1)
- Dräger — Standalone Infinity M540 patient monitor (lower than VG4.1.1)
- Dräger — Standalone Infinity M540 patient monitor (VG4.2)
- Dräger — Infinity Acute Care System (0)
- Dräger — Standalone Infinity M540 patient monitor (0)

## References

- [CNA](https://static.draeger.com/security)
- [CNA](https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling)
- [CNA](https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf)

## Exploitation Prediction (EPSS)

- **EPSS Score:** 0.13%
- **EPSS Percentile:** 3.1

---